Instructions for Use

Cervico Premium Kit

The Cervico Guide with the Premium Mold and off-centred insert: depth and orientation control and duplicate impression posts.

Document details VPI-IFU-PK · rev 2.0
REF
5213006650024
Document
VPI-IFU-PK
Revision
2.0
Date of issue
2026-09-09
CEClass IRx onlyUSANon-sterilesterilize before useMax145 °C

In the box

Cervico Guide

  • Silicone case
  • 12 anatomical tabs (a, p, m1, m2 × S, M, L)
  • 8 cylindrical tabs 5–12 mm
  • 4 guide pins
  • Magnetic retention handle

Cervico Premium Mold

  • Rotating mold base (top ring with codes, bottom ring with 36 sockets)
  • Centered silicone insert
  • Screwdriver
  • VPI Prosthetic Connection Insert(s) as ordered

Cervico Guide in 6 steps

  1. Clean and sterilize first

    Non-sterile on delivery. Ultrasonic clean 10 min at 50 °C, rinse, dry, pouch, autoclave 4 min at 134 °C, dry 30 min.

  2. Tie floss through every part that goes in the mouth

    Tabs and pins have a bore for it. No exceptions.

  3. Pick the tooth group, start with M

    Mount the medium tab, try it in the site, then S or L. Thick biotype: crown width. Thin: one size smaller.

    Cervico Premium Kit — Instructions for Use
  4. Chart the code and the T-line direction

    E.g. "pS, hex flat facing buccal". The code is the mold well; the T-line is where the implant connection reference must face.

  5. Center the osteotomy

    Largest cylindrical tab that touches both neighbours, central line on the occlusal line, pilot drill below 2.1 mm through the bore.

  6. Check every drill with a pin and the tab

    Pin in the osteotomy, tab on the pin, centered or off-centered. Correct at the next drill. Place the implant with its connection facing the T-line.

Cervico Premium Mold in 7 steps

  1. Clean and sterilize first

    Non-sterile on delivery. Remove silicone and inserts, ultrasonic clean 10 min at 50 °C, rinse, dry, pouch, autoclave 4 min at 134 °C, dry 30 min.

  2. Fit your VPI insert once

    Silicone out. Socket group by underline: none = crestal, 1/2/3 lines = 1/2/3 mm subcrestal. Arrow to the flat, screw from below. Write socket and insert on the info card. Silicone back in.

    Cervico Premium Kit — Instructions for Use
  3. Turn the ring to the charted well

    Well code from the Guide (e.g. pS) over your insert. Assistant task.

  4. Abutment into the insert

    Screw the temporary abutment into the VPI insert. Orientation and depth are now fixed.

    Cervico Premium Kit — Instructions for Use
  5. Fill and cure

    Flowable nano-hybrid composite; large wells in two increments. 60 s in the well, 60 s out.

    Cervico Premium Kit — Instructions for Use
  6. Out, polish, disinfect

    Lift out. Polish smooth. Disinfect like any abutment before it goes in the mouth.

    Cervico Premium Kit — Instructions for Use
  7. Impression post: same well, same insert

    Stock impression post instead of the abutment gives the 1:1 duplicate for the impression.

    Cervico Premium Kit — Instructions for Use

Never skip

  • Loose part in the mouth: refer immediately, chest X-ray.
  • Max 145 °C. No dry heat, no chemiclave.
  • Not for patients allergic to aluminium or stainless steel.
  • The mold never goes in the mouth; the abutment does, so polish and disinfect it.
  • Wrong socket group = wrong depth. Check the underlines.
  • Max 145 °C. Replace a torn silicone insert.

After every patient

  • Wipe wells with alcohol, take the silicone out of the base, tabs and pins off the handle.
  • Ultrasonic clean 10 min at 50 °C, aldehyde-, phenol- and chlorine-free detergent. Rinse with purified water. Dry.
  • Inspect. Pouch. Autoclave: 4 min at 132–134 °C pre-vacuum or 30 min at 121 °C gravity. Dry 30 min.
  • Never above 145 °C. Never dry heat, chemiclave, EtO or plasma.

Full instructions

This card does not replace the Instructions for Use. Read VPI-IFU-PK before first use: https://vpicervico.com/instructions-of-use/

VP INNOVATO HOLDINGS LTD · Poseidonos 4A, Aradippou 7101, Larnaca, Cyprus · +357 99 883 898 · info@vpicervico.com

REF 5213006650024 · Class I medical device, CE · Rx only (USA) · Non-sterile, sterilize before use · VPI-QSG-PK rev 2.0, 2026-09-09

VP INNOVATO HOLDINGS LTD · Poseidonos 4A, Aradippou 7101, Larnaca, Cyprus · +357 99 883 898 · info@vpicervico.com
VPI-QSG-PK revision 2.0, issued 2026-09-09. English (original). Latest version: https://vpicervico.com/instructions-of-use/