Instructions for Use

Cervico Premium Mold

Anatomical mold with VPI prosthetic connection inserts for depth- and orientation-controlled custom healing abutments and 1:1 duplicate impression posts.

Document details VPI-IFU-PM · rev 2.0
REF
5213006650000
Document
VPI-IFU-PM
Revision
2.0
Date of issue
2026-09-09
CEClass IRx onlyUSANon-sterilesterilize before useMax145 °C

In the box

  • Rotating mold base (top ring with codes, bottom ring with 36 sockets)
  • Centered silicone insert
  • Screwdriver
  • VPI Prosthetic Connection Insert(s) as ordered

7 steps

  1. Clean and sterilize first

    Non-sterile on delivery. Remove silicone and inserts, ultrasonic clean 10 min at 50 °C, rinse, dry, pouch, autoclave 4 min at 134 °C, dry 30 min.

  2. Fit your VPI insert once

    Silicone out. Socket group by underline: none = crestal, 1/2/3 lines = 1/2/3 mm subcrestal. Arrow to the flat, screw from below. Write socket and insert on the info card. Silicone back in.

    Cervico Premium Mold — Instructions for Use
  3. Turn the ring to the charted well

    Well code from the Guide (e.g. pS) over your insert. Assistant task.

  4. Abutment into the insert

    Screw the temporary abutment into the VPI insert. Orientation and depth are now fixed.

    Cervico Premium Mold — Instructions for Use
  5. Fill and cure

    Flowable nano-hybrid composite; large wells in two increments. 60 s in the well, 60 s out.

    Cervico Premium Mold — Instructions for Use
  6. Out, polish, disinfect

    Lift out. Polish smooth. Disinfect like any abutment before it goes in the mouth.

    Cervico Premium Mold — Instructions for Use
  7. Impression post: same well, same insert

    Stock impression post instead of the abutment gives the 1:1 duplicate for the impression.

    Cervico Premium Mold — Instructions for Use

Never skip

  • The mold never goes in the mouth; the abutment does, so polish and disinfect it.
  • Wrong socket group = wrong depth. Check the underlines.
  • Max 145 °C. Replace a torn silicone insert.

After every patient

  • Wipe wells with alcohol, take the silicone out of the base, tabs and pins off the handle.
  • Ultrasonic clean 10 min at 50 °C, aldehyde-, phenol- and chlorine-free detergent. Rinse with purified water. Dry.
  • Inspect. Pouch. Autoclave: 4 min at 132–134 °C pre-vacuum or 30 min at 121 °C gravity. Dry 30 min.
  • Never above 145 °C. Never dry heat, chemiclave, EtO or plasma.

Full instructions

This card does not replace the Instructions for Use. Read VPI-IFU-PM before first use: https://vpicervico.com/instructions-of-use/

VP INNOVATO HOLDINGS LTD · Poseidonos 4A, Aradippou 7101, Larnaca, Cyprus · +357 99 883 898 · info@vpicervico.com

REF 5213006650000 · Class I medical device, CE · Rx only (USA) · Non-sterile, sterilize before use · VPI-QSG-PM rev 2.0, 2026-09-09